Home » Product List » Blood Analyzer Equipment » XFA6000/6000A/6000B Auto Hematology Analyzer
Product Details
XFA6000/6000A/6000B Auto Hematology Analyzer
| Packing: | carton |
|---|---|
| Model NO.: | XFA6000 |
| Standard: | ISO9001, ISO13485, TuV |
| Productivity: | 50, 000 pcs/year |
| Trademark: | Perlong |
| Origin: | Nanjing, China |
| Min. Order: | 1 set |
| Transportation: | by sea, air, railway etc |
Product Description
Auto Hematology Analyzer(3 Difference 22 Parameters)
Test Items WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW_CV, RDW_SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram
Background Count WBC: ≤ 0.3× 109/L RBC: ≤ 0.03× 1012/L HGB: ≤ 1g/L PLT: ≤ 10× 109/L
| Test Items | WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram | ||
| Background Count | WBC: ≤0.3×10 9 /L RBC: ≤0.03×10 12 /L HGB: ≤1g/L PLT: ≤10×10 9 /L | ||
| Linearity | Test Item | Test Range | Allow Difference of Linearity |
| WBC | 0∼6.0×10 9 /L | ±0.3×10 9 /L | |
| 6.0∼99.9×10 9 /L | ±5% | ||
| RBC | 0∼0.99×10 12 /L | ±0.05×10 12 / L | |
| 1.0∼9.99×10 12 /L | ±5% | ||
| HGB | 0∼99 g/L | ±2g/L | |
| 100∼300 g/L | ±2% | ||
| PLT | 0∼99×10 9 /L | ±10×10 9 /L | |
| 100∼999×10 9 /L | ±10% | ||
| Repeatability | Test Item | CV(%) | Teat Range |
| WBC | ≤2.0 | 7.0~15.0×10 9 /L | |
| RBC | ≤0.5 | 3.5~6.0×10 12 /L | |
| HGB | ≤1.5 | 110~180g/L | |
| PLT | ≤4.0 | 150~500×10 9 /L | |
| MCV | ≤0.5 | 80~110fL | |
| Sample Volume | Twig blood: 20μl | ||
| Dilution Ratio | WBC/HGB: 1:500 RBC/PLT: 1:50000 | ||
| Interfaces | RS232 joint, USB, PS2, Parallel port | ||
| Test Speed | 60 samples/hours | ||
| Continuous Work Time | >8h | ||
Inquiry Basket (